Erratum: High Flow Repatriation – will we be running out of air while in the air?

ITIJ Air Ambulance Review March 2023
In the Case Report “High Flow Repatriation – will we be running out of air while in the air?”, published in ITIJ Air Ambulance Review March 2023, the formula to calculate the oxygen consumption of the ventilator during HFNC and the values derived from that formula are incorrect. The correct formula should read (FiO2 - 0.21) x Flow rate [l/min] / 0.79 = oxygen flow per minute.
A corrected version of the paper can be found online here.